CSInstruments HD-KFM III: KPFM surface potential on the cantilever's second eigenmode, with topography, the tip 0.1–0.5 nm from the surface.
Key features
- Second eigenmode
- Surface potential detected on the cantilever's second eigenmode, at 350–400 kHz, while the first mode follows the topography: both channels in one scan.
- Tip close to the surface
- The tip stays 0.1–0.5 nm from the surface, instead of 10–100 nm in lift-mode KPFM.
- Low noise
- Signal amplified by the Q factor of the second eigenmode; noise level below 1 mV with a small AC voltage.
- Below the surface
- Reveals structures buried under the surface, such as nanotubes in a polymer.
- More channels
- Capacitance gradient dC/dZ at the same time, and MFM with electric field compensation (EFC) in the same scan.
- Runs on
- The Nano-Observer II. On a Bruker MultiMode or a Keysight / Agilent 5100 or 5500, the Galaxy Dual Controller adds it.
Specifications
From Concept Scientific Instruments' product page.
| Potential detection | 350–400 kHz (second eigenmode) |
|---|---|
| Noise level | < 1 mV |
| Tip–sample distance | 0.1–0.5 nm |
| Lateral resolution | Below 10 nm (lift-mode KPFM: typically 20–50 nm) |
| Simultaneous channels | Topography, surface potential, dC/dZ; MFM + KFM in the same scan |
Overview
HD-KFM III is the Kelvin probe force microscopy (KPFM) mode for the CSInstruments Nano-Observer II AFM. HD-KFM measures surface potential on the second eigenmode of the cantilever, at the same time as topography, with higher resolution and sensitivity than lift-mode KFM.
Standard KPFM reads the potential in a second pass, with the tip lifted 10–50 nm above the surface. HD-KFM III drives the cantilever’s second eigenmode electrically, at 350–400 kHz, 20 to 30 times higher than side-band KPFM. The higher Q factor of that mode gives a low-noise signal with a smaller AC voltage.
It is used wherever a difference in surface potential or work function matters: layer counting and moiré domains in 2D materials, doping contrast and junctions in semiconductors, grain boundaries in perovskite solar cells and charge distribution in polymers.
Webinars and videos
From the manufacturer's own channels.
-